- Contact Person : Ms. Shi Helen
- Company Name : Beijing Xin He Feng Medical Technology Co., Ltd.
- Tel : 0086-10-67972240
- Address : Beijing,Beijing,No.19 Tianrong Street,Biomedical Industrial Base,Daxing District,Beijing,China.
- Country/Region : China
- Zip : 102609
Related Product Searches:FC112 battery syringe pump,High Quality,battery syringe pump, channel syringe pump,EA1
Model No. Parameters | FC112 | |
Wi-Fi | Yes | |
Dru Library | Yes | |
Infusion Modes | Rate Mode,Time Mode,Body-Weight Mode | |
Drug Name Display | More than 50,can be updated | |
Safety | Double CPU to make sure higher standard infusion | |
Accuracy | ±2% | |
Time Preset | 0~99h and 59min | |
Volume Range | 0.01~9999.99ml | |
Flow Rate Step by | 0.01 ml/h | |
Flow rate range | 5ml Syringe 0.01ml/h-150ml/h | |
10ml Syringe 0.01ml/h-400ml/h | ||
20ml Syringe 0.01ml/h-600ml/h | ||
30ml Syringe 0.01ml/h-900ml/h | ||
50/60ml Syringe 0.01ml/h-1500ml/h | ||
Purge/Bolus rate/Top flow rate | 5ml Syringe 150ml/h | |
10ml Syringe 400ml/h | ||
20ml Syringe 600ml/h | ||
30ml Syringe 900ml/h | ||
50/60ml Syringe 1500ml/h | ||
Audible & Visible Alarms | Syringe Dislocation,Occlusion,Near End,Pause Overtime,Low Battery,Battery Exausted,VTBI Completion,Syringe Empty,Motor Malfunction,Incorrect Syringe Specification,Circuit Malfunction,Internal Communication Malfunction,Master CPU Malfunction,Monitoring CPU Malfunction | |
KVO | Rate≥10ml/h:KVO=3ml/h | |
1ml/h≤Rate≤10ml/h:KVO=1ml/h | ||
Rate ≤1ml/h:KVO=Preset rate(Defaulted rate: 0.1 ml/h) | ||
Occlusion Pressure | High: | 106.7KPa±26.7 KPa |
Middle: | 66.7KPa±13.3 KPa | |
Low: | 40.7KPa±13.3 KPa | |
Pause Overtime Alarm | 1min50s—2min | |
Power & Battery | 25VA, AC:100V-240V 50HZ/60HZ, DC:12V,Lithium battery support more than 8 hours | |
Classification | Class II,type CF,IPX4 |
FC112 battery syringe pump